Wireless temperature/humidity/pressure verification system

DT Pro Wireless Temperature / Humidity / Pressure Calibration System

2021-11-25 This article is read 7029 Times

Wireless Temperature / Humidity / Pressure Calibration System

◇ Introduction ◇ Function◇ Parameter◇ Product order◇ Download◇ Related products

Introduction

DT Pro Wireless Temperature / Humidity / Pressure Calibration System can be widely used in equipment calibration and environmental monitoring in pharmaceutical, food, biological, chemical and other industries. Wireless calibration system Compared with the traditional wired calibration system, the probe can be conveniently and quickly arranged, and the detection point is physical constraints greatly improve calibration efficiency.

Function

The system consists of a wireless recorder (temperature, humidity, pressure), a data reading workstation, and wireless calibration software.

DT Pro temperature calibration system can effectively store 16000 pieces of data, and the temperature measurement accuracy is 0.1 ℃, which fully meets the needs of equipment and process calibration in food and drug production enterprises and the calibration needs in the fields of bioengineering, medical and transportation. The recorder uses replaceable batteries, with large power, low power consumption and long battery life. Adopt high pressure and waterproof design, can be used continuously for a long time in high pressure and high humidity environment.

The data reading workstation is divided into a single-hole workstation and a 10-hole workstation. The universal USB serial port is used to connect to the computer, and data communication can be carried out conveniently, efficiently and safely.

The software realizes data collection, data analysis and report generation, meets FDA 21 CFR Part 11, and supports electronic signature of documents. Support custom stages and verify each stage to generate stage calibration report.

Features:

◇ Switch between Chinese and English versions at will;

◇ Comply with FDA 21CFRPart 11 regulations;

◇ With data audit tracking function, with three levels of authority (freely assignable authority);

◇ It has data statistical analysis functions, including large and small values, average value, span, temperature difference at the same time, F0 value test, etc., and can be freely selected according to needs.

Temperature Calibration System

Application scenarios:

DT Pro Pressure / Temperature / Humidity Calibration System can be applied to GMP pharmaceutical industry, GSP industry, biological industry, medical industry, food industry, transportation industry.

GMPSteam sterilizer calibration / water bath sterilizer verification, rubber plug washer calibration / aluminum lid washer verification, freeze dryer calibration / cryogenic refrigerator verification, constant temperature and humidity chamber calibration / stability incubator calibration
GSPCold storage verification / refrigerated truck calibration, refrigerator / incubator verification, cool storage / normal temperature storage calibration
BiologicalStirred fermentation tank verification, culture bath calibration, process development data
FoodPasteurization verification, sterilization tunnel calibration, rotary freezer verification, blast freezer calibration, cold room calibration
TransportCold chain process tracking, cabin temperature detection
MedicalHigh temperature sterilization testing, steam sterilization testing, medical supplies testing

Implemented procedures and specifications:

Serial numberRegulation codeProcedure name
1GBT 30690-2014The sterilization effect monitoring method and evaluation requirements of small pressure steam sterilizer
2PDA TR1-2007Damp heat sterilization verification
3JJF 1366-2012"Calibration Regulations for Temperature Data Collector"
4JJF 1101-2003"Code for Calibration of Temperature and Humidity of Environmental Test Equipment"
Related regulations"Pharmaceutical Production Quality Management Standards" (revised in 2015) "Guidelines for Pharmaceutical Production calibration" (2003 edition) FDA 21CFRPart11 Clause EN285-2006 (German Standard) EN554: 1994 (German Standard) HTM2010 (British Standard)

Technical Parameters


Temperature CalibratorTemperature and humidity calibratorPressure Calibrator
ModelTDLATDLC(Flexible probe)MDLAPDLA
Operating temperature-50 ~ 150 ° C (expandable to 400?C with heat insulation box)-50 ~ 150 ° C (expandable to 400?C with heat insulation box)-40~125°C-40~140°C
Range-50~150°C-50~150°C0~100%RH0~100psi
0~6.894Bar
(Absolute pressure)
Resolution0.01°C0.01°C0.01°C0.008psi;
0.01%RH0.55mBar
Accuracy±0.1°C±0.1°C±0.2°C±2.5% Full scale
±2%RH
SensorPt100 Class 1/3 DINPt100 Class 1/3 DINTemperature: Pt100Strain
Humidity: capacitive
Capacity (recording points)64,00064,0008,0008,000
Record frequency1 second to 18 hours1 second to 18 hours1 second to 18 hours1 second to 18 hours
Battery Life36 months36 months12 months12 months
Communication methodRS 485 ContactRS 485 ContactRS 485 ContactRS 485 Contact
DT Pro Wireless Temperature / Humidity / Pressure Calibration System
  • Online message

  • Your name
    Your title must be filled
  • Cellphone number
    Mobile phone number is required
  • whatsapp/wechat
    Optional
  • Type of consultation
    Required, please select the type of consultation
  • Consultation content
    Required, what is the content you want to consult?
  • Verification code: Please fill in the calculation result.
Previous:DTWX-01G Wireless Temperature Humidity Acquisition/Inspection Systemnext:Custom temperature and humidity verification/calibration chamber

Copyright? Tai'an Dearto Automation Instrument Co., Ltd.

Chinese