0538-5089056
◇ Introduction ◇ Function◇ Parameter◇ Product order◇ Download◇ Related products
Introduction:
Class I Standard Platinum Rhodium 10-Platinum Thermocouple (S) is a temperature measurement standard,has standard platinum-rhodium 10-platinum thermocouple and standard platinum-rhodium 30-platinum-rhodium 6 thermocouple two categories. Standard platinum-rhodium 10-platinum thermocouple is used in the temperature range of 419.527~1084.62℃ to achieve the value of thermocouple temperature transfer and precision temperature, accuracy grade has first-class standard and second-class standard. Standard platinum-rhodium 30-platinum-rhodium 6 thermocouple used in 1100~1500℃ temperature area thermocouple temperature value transfer and precision temperature, precision grade has second standard.
Technical indicator:
Type | Standard platinum rhodium10-Platinum | Standard platinum rhodium30-platinum rhodium 6 | |
Temp.range | 300~1300℃ | 1100~1500℃ | |
Model | WRPB-1 | WRPB-2 | WRRB-2 |
Accuracy grade | Grade I | Grade II | Grade II |
Stability | Better than 3μV | Better than 5μV | Better than 8μV |
Thermoelectric range | E(tCu)=10.575±0.015mV E(tAl)=5.860+0.37〔E(tCu)-10.575〕±0.005mV E(tZn)= 3.447+0.18〔E(tCu)-10.575〕±0.005mV | E(1100℃)=5.780±0.025mV E(1500℃)=10.099±0.040mV
| |
Electrode Size
| DiameterΦ0.5 Length not less than1000mm |
Previous:Class II Standard Platinum Resistance(Zinc Point Quartz Tube) | next:Custom temperature and humidity verification/calibration chamber |
Headquarter Address :Tai'an High-tech Development Zone, Shandong Province
Telephon :+86 0538-5089056 +86 13605384645 +86 13953856217
Fax : +86 0538-5059718
After sale:+86 0538-5050959 +86 13605384645
Website : www.tadt.com.cn
E-mail : tadt17@dearto.cn / tadtzdh@dearto.cn
Copyright? Tai'an Detu Automation Instrument Co., Ltd. Lu ICP 14024757-1